BRUEL & KJAER 2032 Instruction Manual page 79

Dual channel signal analyzer
Table of Contents

Advertisement

b)
Frequency Domain Functions
Up to six Y -scaling control fields are used to display a frequency domain function.
They control:
1.
Y -axis full scale
2.
Y -axis units, absolute or relative
3.
Y -axis dB reference
4.
Y -axis spectrum units
5.
Y -axis range in dB
6.
Y -axis lin.-log.
Field 3 disappears from the 2032 Display when read-out in absolute units is selected,
and Field 5 disappears when a linear Y -axis is selected. Note that while read-out with
logarithmic scaling is possible in both absolute and relative units, read-out with linear
scaling is only possible in absolute units.
v:CJ
~9E9l
r
or
100E~15
V
2
or U
2
*
0.
2
2
V /Hz or U /Hz*
V
2
s/Hz or U
2
s/Hz*
~9E9l
r
or
u•
:
V
2
or U
2
*
.
0
·
·
V
2
/Hz or U
2
/Hz*
100E:_15
V
2
s/Hz or U
2
s/Hz*
500
499.9
-499.9
-500
1.d8
D
(dB***
ref.
level)
0. RMS
l
1. PWR
2. PSD
3.
ESD
0. RMS
l
1. PWR
2. PSD
3.
ESD
l
0.
RMS
l
1. PWR
2. PSD
3.
ESD
D
10
20
40
80
160
10
20
40
80
160
D
0. LIN**
1.
dB
1.
dB
• V, etc., when the 2032 Is calibrated In volts, U, etc., when the 2032 is calibrated in engineering
units.
• • Not allowed for Nichols plots.
• • • For entry of the user defined reference level, which should be in units appropriate to the spectrum
units
field.
Allowed range
.
999 E 9 to 100 E-15. Should the spectrum units field be changed, the
reference level will automatically be changed as
appropriate.
Note that reference levels entered
when the spectrum units field is set to RMS can achieve the range 998 E 21 to 10.0 E-27 should
display of PWR, PSD, or ESD be selected.
69