RECORD:
r - - - - - R e c o r d - - - - - .
Cont
.
Single
Stop
I l l
MANUAL TRIGGER:
Manual
Trigger
I
ll
MEMORY:
r -
Memory -----,
Enable
Store
II
~
m
30
a
"Stop", or to repeat a linear or peak averaging cycle, averaging the
results with those already obtained.
If the 2032 averager is in Manual Accept mode, "Proceed" causes the
instantaneous function(s) present in the 2032 at that time to be en-
tered into the average.
"Stop" -
stops the averaging process.
"Cont." -
causes new records to be continously taken and pro-
cessed according to the triggering, delay, and overlap conditions set.
"Single" -
holds the first record satisfying the delay and triggering
conditions set, switching automatically to RECORD "Stop". The re-
cord is then processed.
"Stop" -
immediately stops the recording process. Setting the 2032
to RECORD "Single" and letting it automatically switch to RECORD
"Stop" ensures that the ongoing record is completed.
Allows the manual capture of signals, with full use of the trigger delay
to the start of channel A or 8 and delay between channel A and 8
controls.
The MANUAL TRIGGER pushkey can be used in all trigger modes.
"Enable" -
enables "Store".
"Store" -
causes the 2032 to store the measured functions in mem-
ory if "Enable" is active, and then disables "Enable", hence protect-
ing the memory contents. The memory can only be further updated by
again pressing "Enable" and "Store".
The stored functions are as follows:
Single Spectrum Averaging Ch. A or 8
2048 sample Time Function
1024 line Autospectrum
Single Spectrum Averaging Zero Pad Ch. A or 8
1024 sample Time Function
1024 line Autospectrum
Dual Spectrum Averaging
2 x 2048 sample Time Functions
2 x 1024 line Autospectra
1 x 1024 line Cross Spectrum
Dual Spectrum Averaging Zero Pad
2 x 1024 sample Time Functions
2 x 1024 line Autospectra
1 x 1024 line Cross Spectrum