BRUEL & KJAER 2032 Instruction Manual page 144

Dual channel signal analyzer
Table of Contents

Advertisement

DIGITAL 1/0
80 MANUAL DIG OUT .................................................................................................... 12.7.1
81 MAN DIG OUT PAR 1 ............................................................................................... 12.7.2
82 MAN DIG OUT PAR 2 ............................................................................................... 12.7.3
83 MAN DIG OUT PAR 3 ............................................................................................... 12.7.4
84 REPEAT DIG OUTPUT ..............................................................................................
12~
7.5
86 DISPL 1/0 STATUS ................................................................................................... 12.7.6
87 CR AND TERMINATOR ............................................................................................ 12.7.7
SPL-EXECUTION
90 SPL SOURCE
91 DIG TEST SIGNAL
92 SPL ENTRY ADDRESS
93 SPL BREAK ADDRESS
95 SPL STEP /TRAP KEY
96 SPL STEP
97 SPL TRAP ADDRESS
SERVICE
100 ZOOM KEY
101 ZOOM PAR 0
102 ZOOM PAR 1
103 ZOOM PAR 2
104 ANALOG INPUT KEY
105 ANALOG INP PAR 0
106 ANALOG INP PAR 1
107 ANALOG INP PAR 2
108 ANALOG INP PAR 3
109 DISPL KEY CODE
See 2032 SPL Manual
See the 2032 Service Instructions
12.4.1. 0: Main Key
Main Key acts as a key allowing entry to the other Special Parameters, the Main
Memory, the Scratch Pad Memory, and the Continuous Memory. Entry is only allowed
when Main Key is set to 2032 or 2034, see Section 12.3.3.
12.5. SPECIAL PARAMETERS, MEASUREMENT
12.5.1. 1: Trigger Hysteresis
134
Trigger Hysteresis sets the minimum change in input signal level allowed using internal
trigger, that is, Ch.A, Ch.A filtered, Ch.B, or Ch.B filtered, before a trigger can occur. It
is set in percentage of the peak input voltage setting of the trigger channel, and the
default value is 5%.