Patient connections

Selecting the sensor

GE Healthcare OxyTip+ Finger Sensors are recommended for short term patient monitoring
with patients weighing more than 20 kg.
The EarSat sensors are recommended for patients weighing more than 10 kg and patients with
compromised peripheral circulation.
For long term monitoring and during high motion conditions, use GE Healthcare OxyTip+
Adhesive Sensors.
For pediatric patients weighing more than 3 kg, use GE Healthcare OxyTip+ Adhesive Sensors
or Wrap Sensors.
NOTE: For each SpO
package for patient weight limits.
1
Figure 12-2 Pulse oximetry setup with OxyTip+ sensors
(1)
E-PSMW or E-PSMPW module with SpO
(2)
OxyTip+ interconnect cable
NOTE: The listed sensors are latex-free.
accessory, refer to the instructions for use in the accessory
2
2
2
OXY-AP
measurement capability
Pulse oximetry
OXY-E-UN
OXY-SE-3
OXY-F-UN
OXY-W-UN
OXY-AF
12-3