P R E L I M I N A R Y
DIRECTION 5854713-100, REVISION 2
3-5-3

Electrical Specifications

WARNING
WARNING
3-5-3-1
Verification of the Venue Fit™ Voltage Setting
Verify that the mains voltage specified for the Venue Fit™ is available on-site.
The voltage setting for the Venue Fit™ is found on a label near the Mains Power Circuit Breaker on the
rear of the Venue Fit™ .
3-5-3-2

Electrical Specifications for the Venue Fit™

In the table below, the electrical specifications for Venue Fit™ includes monitor and on board
peripherals.
Table 3-15
The system only current drain will vary depending on the mains voltage.
At 240 VAC the current may be up to 1.04 A.
At 100 VAC the current may be up to 2.5 A.
3-5-4

Connections on the I/O Rear Panel

NOTE:
Accessory equipment connected to the analog and digital interfaces must be certified according to the
respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore, all complete configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everybody who connects additional equipment to the signal input part or signal
output part of Venue Fit™ , configures a medical system, and is therefore responsible that the
Ultrasound system complies with the requirements of the valid version of IEC60601-1-1. If in doubt,
consult the technical service department or your local representative for GE.
3-5-4-1

Network Connection

Connect the network cable to the Ethernet connector on the External I/O.
The connector is located on the rear side of Venue Fit™ .
3-5-5

Connecting Probes

3-5-5-1

Introduction to Connecting Probes

Probes can be connected or changed any time, as described below, regardless of whether the system
is powered ON or OFF.
3-16
Connecting a Venue Fit™ to the wrong voltage level will most
likely destroy it.
Electrical Specifications for all Venue Fit™ Models
Voltage
Power Consumption
(system only)
Power Consumption
(with cart and peripherals)
Frequency
Chapter 3 - System Setup
100-240 VAC
±10%
250 Watt
350 Watt
50-60 Hz
VENUE FIT™ SERVICE MANUAL