3-6-6 Connections on the I/O Rear Panel

NOTE:
3-6-6-1 Connect Ethernet
3-6-6-2 Connect USB Flash Drive
NOTE:

3-6-7 Connecting probes

3-6-8 Powering the system

3-24
System view
Accessory equipment connected to the analog and digital
interfaces must be certified according to the respective IEC
standards (e.g. IEC60950 for data processing equipment and
IEC60601-1 for medical equipment). Furthermore, all complete
configurations shall comply with the valid version of the system
standard IEC60601-1. Everybody who connects additional
equipment to the signal input part or signal output part of
Versana Balance, configures a medical system, and is therefore
responsible that the Ultrasound system complies with the
requirements of the valid version of IEC60601-1. If in doubt,
consult the technical service department or your local
representative for GE.
Connect the network cable to the Ethernet connector on the I/O
Rear Panel.
The connector is located on the rear side of Versana Balance.
USB Flash Drive approved for Versana Balance are verified for
EMC performance according to EN55011 class B. The use of
any other USB Flash Drive will compromise this verification, and
may cause interference on Versana Balance itself, or on other
electronic devices.
For approved models, please refer to Chapter 9.
Insert the USB Flash Drive in one of the USB ports on the
Versana Balance.
Please refer to Basic User Manual on how to connect/
disconnect a probe.
Please refer to Basic User Manual on how to power the system.
Versana Balance – Basic Service Manual
5808768-100 English Rev.9