Electrical Specifications; Connections On The I/O Rear Panel; Connecting Probes - GE Venue Go Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5813707-100, R
IRECTION
3-5-3

Electrical Specifications

WARNING
WARNING
3-5-3-1
Verification of the Venue Go™ Voltage Setting
Verify that the mains voltage specified for the Venue Go™ is available on-site.
The voltage setting for the Venue Go™ is found on a label near the Mains Power Circuit Breaker on
the rear of the Venue Go™ .
3-5-3-2
Electrical Specifications for the Venue Go™
In the table below, the electrical specifications for Venue Go™ includes monitor and on board
peripherals.
Table 3-16
The system only current drain will vary depending on the mains voltage.
At 240 VAC the current may be up to 1.04 A.
At 100 VAC the current may be up to 2.5 A.
3-5-4

Connections on the I/O Rear Panel

NOTE:
Accessory equipment connected to the analog and digital interfaces must be certified according to the
respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore, all complete configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everybody who connects additional equipment to the signal input part or signal
output part of Venue Go™ , configures a medical system, and is therefore responsible that the
Ultrasound system complies with the requirements of the valid version of IEC60601-1-1. If in doubt,
consult the technical service department or your local representative for GE.
3-5-4-1
Network Connection
Connect the network cable to the Ethernet connector on the External I/O.
The connector is located on the rear side of Venue Go™ .
3-5-5

Connecting Probes

3-5-5-1
Introduction to Connecting Probes
Probes can be connected or changed any time, as described below, regardless of whether the system
is powered ON or OFF.
3-16
2
EVISION
Connecting a Venue Go™ to the wrong voltage level will most
likely destroy it.
Electrical Specifications for all Venue Go™ Models
Voltage
Power Consumption
(system only)
Power Consumption
(with cart and peripherals)
Frequency
Chapter 3 - System Setup
100-240 VAC
±10%
250 Watt
350 Watt
50-60 Hz
V
G
™ S
M
ENUE
O
ERVICE
ANUAL

Advertisement

Table of Contents
loading

This manual is also suitable for:

H45601vg