P R E L I M I N A R Y
D
5813707-100, R
IRECTION
3-5-3
Electrical Specifications
WARNING
WARNING
3-5-3-1
Verification of the Venue Go™ Voltage Setting
Verify that the mains voltage specified for the Venue Go™ is available on-site.
The voltage setting for the Venue Go™ is found on a label near the Mains Power Circuit Breaker on
the rear of the Venue Go™ .
3-5-3-2
Electrical Specifications for the Venue Go™
In the table below, the electrical specifications for Venue Go™ includes monitor and on board
peripherals.
Table 3-16
The system only current drain will vary depending on the mains voltage.
•
At 240 VAC the current may be up to 1.04 A.
•
At 100 VAC the current may be up to 2.5 A.
3-5-4
Connections on the I/O Rear Panel
NOTE:
Accessory equipment connected to the analog and digital interfaces must be certified according to the
respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore, all complete configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everybody who connects additional equipment to the signal input part or signal
output part of Venue Go™ , configures a medical system, and is therefore responsible that the
Ultrasound system complies with the requirements of the valid version of IEC60601-1-1. If in doubt,
consult the technical service department or your local representative for GE.
3-5-4-1
Network Connection
Connect the network cable to the Ethernet connector on the External I/O.
The connector is located on the rear side of Venue Go™ .
3-5-5
Connecting Probes
3-5-5-1
Introduction to Connecting Probes
Probes can be connected or changed any time, as described below, regardless of whether the system
is powered ON or OFF.
3-16
2
EVISION
Connecting a Venue Go™ to the wrong voltage level will most
likely destroy it.
Electrical Specifications for all Venue Go™ Models
Voltage
Power Consumption
(system only)
Power Consumption
(with cart and peripherals)
Frequency
Chapter 3 - System Setup
100-240 VAC
±10%
250 Watt
350 Watt
50-60 Hz
V
G
™ S
M
ENUE
O
ERVICE
ANUAL