GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-14

Keycap set, Main Keyboard Encoder Knob set

8-2-14-1
Tools
Common pilIips screwdrivers
8-2-14-2
Preparations
Shut down the system and switch off the main breaker.
8-2-14-3
Removal procedure
1.) Remove key caps from the keyboard.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-15
Service Manual
Section
Section 4-3-1
Section 4-3-2
Section 4-8-4
Section 10-5-5
8-2-14-4
Mounting Procedure
Install the new parts in the reverse order of removal.
8-50
5
Figure 8-83 Key Caps
Functional Tests
Functional Test / Diagnostic Test
Power On/Boot Up
Power Off / Shutdown
TGC key assy/TGC Knob Set, Sub keyboard
encoder Knob Set, Keycap set, Main Keyboard
Encoder Knob set Function Validation
Physical Inspection
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
5394141, Rev 1+, Section 8-2-14. Equipment
passed all required tests and is ready for use. "
Procedure
LOGIQ™ P5 S
M
EVICE
ANUAL
Debrief Script
"Service Manual, Direction