Circuit O2 Cell Replacement - GE Carestation 620 User's Reference Manual

Table of Contents

Advertisement

Carestation
620/650/650c (A1)

Circuit O2 cell replacement

WARNING
9-6
Handle and dispose of O2 cells according to site
biohazard policies. Do not incinerate.
Note
It may take a new O2 cell 90 minutes to stabilize. If the O2 cell
calibration fails after a new O2 cell is installed, wait 90 minutes and
repeat the calibration.
Note
Make sure O2 cell plug is removed if adding an O2 cell to the system
(see section "Circuit O2 cell installation").
1.
Pull down the canister lifter handle to unlock the canister from
the breathing system.
2.
Lift the canister and remove it from the canister lifter base.
3.
Open the breathing system door. Push the release button to
unlock the breathing system.
4.
Lower the breathing system.
5.
Pull out the O2 cell. Remove the cable connector from the O2
cell.
2076152-001