Firmware Update

Firmware Update
Firmware Update

Required Equipment

NOTE
The free space of 30 MB or more is necessary for the USB mass storage device.

Location of the latest E5071C Firmware

Step 1. The latest firmware can be downloaded from the following site.

How to update the E5071C firmware

It is possible to update the E5071C firmware, but not the following software by this
procedure.
The following is the procedure to update the E5071C firmware.
NOTE
User should log in as "agena", and user should have administrator authority when
performing firmware update.
Step 1. Prepare the latest E5071C firmware (USB mass storage device).
Step 2. Connect the mouse and the keyboard to the E5071C rear panel. Then turn the E5071C on.
Step 3. Connect the USB mass storage device into the front USB port of the E5071C.
Step 4. Update the E5071C firmware.
1. Press
2. Press
3. Enter the password
Figure C-1
Passward box
e5071cse1081
362
E5071C (with keyboard and mouse).
USB memory.
http://www.agilent.com/find/ena/e5071c/
OS (Operating System)
VBA (Visual Basic for Application)
Calibration Constant Data
key.
-
Service Menu
Update Firmware
will appear.
e507xa
on the menu bar, and the Password dialog box
into the Password box, then click
.
OK
Appendix C