Installation
Always connect the equipment to a dedicated (separate) circuit. Electrical
codes require fuse or circuit breaker protection for branch circuit
conductors. Use time delay fuses for #12 AWG circuits.
For all blood bank models, the wiring diagram is attached to the back of the
cabinet.

7.3 Leveling

CAUTION: An unlevel unit may result in instability and performance
issues for the doors and drawers.
The unit must be level. Rotate the leveling screws, located under the front
corners of the unit, until the unit is level. If the floor is seriously out of level,
you may need to shim the corners with thin sheets of metal.
Note Use thin sheets of metal to level units equipped with optional casters.

7.4 Door Operation

Standard glass doors for models larger than the undercounter model stay
open if opened 90 degrees. Door spring tension cannot be adjusted.
If the self-closing doors do not work properly, make sure the unit is level.
7.4.1 Adjustable Hinged Glass
On Undercounter models with hinged glass doors, you can use a regular
Doors (Undercounter Models
screwdriver to adjust spring tension and center each door on its frame. The
Only)
adjustment screws ("torque" for tension and "sag" for door placement) are
located on the bottom hinge bracket (shown below in Figure 1).
Hinged glass doors can be propped open with metal braces at the bottom
hinges. Open the door 90 degrees or until you feel some resistance. Push the
door open past the resistance and the metal braces engage. To close the door,
push it toward the unit (past the resistance).
Figure 1. Undercounter Hinge Bracket for Glass Door Adjustments