Autoclaving; Service Of Thermo Fisher Scientific - Thermo Scientific Sorvall LYNX 4000 Instruction Manual

Superspeed centrifuge
Hide thumbs Also See for Sorvall LYNX 4000:
Table of Contents

Advertisement

Autoclaving

1. Before autoclaving clean rotor and accessories as described above.
2. Place the rotor on a flat surface.
• Rotors and adapter can be autoclaved at 121 °C.
• The maximum permissible autoclave cycle is 20 minutes at 121 °C.

Service of Thermo Fisher Scientific

Thermo Fisher Scientific recommends having the centrifuge and accessories serviced once a year by an
authorized service technician. The service technician checks the following:
• the electrical equipment;
• the suitability of the set-up site;
• the centrifuge door lock and the safety system;
• the rotor;
• the fixation of the rotor and the drive shaft;
• the protective casing.
Thermo Fisher Scientific offers inspection and service contracts for this work. Any necessary repairs are
performed for free during the warranty period and afterwards for a charge.
This is only valid if the centrifuge has only been maintained by a Thermo Scientific service technician.
Thermo Scientific
CAUTION No chemical additives are permitted in the steam.
Never exceed the permitted temperature and duration when autoclaving.
WARNING If the rotor shows signs of corrosion or wear, it must be replaced.
5