To adjust 10 MHz low pass filter

This procedure adjusts the A35 Analog Source assembly's reconstruction
filter. The reconstruction filter is a 10 MHz low pass filter.
Equipment Required:
Network analyzer
Extender board, Agilent part number 89410-B1001
(2) BNC cables
(2) N(m)-to-BNC(f) adapters
BNC(f)-to-test clips
Flat-edge adjustment tool, Agilent part number 8710-1928
1
2
Place the A35 Analog Source assembly on an extender board.
a
Set the power switch to off ( O ).
b
Remove screws holding the A35 Analog Source assembly in the
card nest. Do not remove the screws in the center of the cover.
c
Place the assembly on an extender board.
d
Set the power switch to on ( l ).
Set up the Agilent 89410A.
a
Press the following keys:
]
[ System Utility
[ more
]
[
]
diagnostics
[
]
service functions
1125
[
enter
]
[
adjustments
]
Adjusting the Analyzer
To adjust 10 MHz low pass filter
2-25