5.7

Setting Up the System in the Software

Loading the USB Drivers
Starting the Instrument Controller and Instrument Configuration Manager
UHPLC System
Operating Manual
This section describes the basic steps for operating the Vanquish system
with the Chromeleon 7 software. For additional information, refer to the
Help and documents provided with the software.
This manual assumes that the Chromeleon software is already installed
on the data system computer and a valid license is available.
TIP In order to start the Chromeleon Instrument Controller, you must log
on under Windows with local administrator privileges, or as a member
of the Windows Chromeleon Operators user group.
1. Turn on the power to the data system computer if it is not yet
already on.
2. Turn on the power to all system modules.
Windows will automatically detect the new devices and perform the
USB installation. If Windows fails to detect the devices and launches
a wizard instead, this indicates that you connected the devices to
the computer and turned on the power for the first time before you
installed the Chromeleon software. In this case, cancel the wizard,
install the software, and then repeat the steps.
1. Start the Chromeleon Instrument Controller.
Right-click the Chromeleon tray icon on the Windows taskbar (the
icon is crossed out in red) and click Start Chromeleon Instrument
Controller. The icon changes to gold, indicating that the Instrument
Controller Service is starting. The icon changes to gray when the
Instrument Controller Service is running (idle).
If the Chromeleon tray icon is not available on the Windows taskbar,
click Start > All Programs (or Programs, depending on the operating
system) >Thermo Chromeleon 7 > Services Manager > Start
Instrument Controller.
2. Start the Chromeleon 7 Instrument Configuration Manager.
Click Start > All Programs (or Programs, depending on the operating
system) > Thermo Chromeleon 7 > Instrument Configuration
Manager.