Charging the battery

4H101 7.5° X3 flexion stop
4H102 15° X3 flexion stop
X3 – flexion stop 22.5°: 4H104
Only when "DUAL" marking is not present (see page 12)
4X350-X3 remote control
The user interface is available in the following languages: German, English, French, Italian, Spanish, Dutch,
Swedish
4X194 wrist band for 4X350-X3 remote control
6 Charging the battery
The following points must be observed when charging the battery:
Use the 757L16* power supply and 4E60* battery charger to charge the battery.
With average use, the capacity of the fully charged battery is sufficient for about 5 days.
We recommend charging the product once a day when used by the patient on a daily basis.
The battery should be charged for at least 3 hours prior to initial use.
Note the permissible temperature range for charging the battery (see page 51).
The distance between the battery charger and the receiver on the product must not exceed 2 mm.
The tube adapter must be connected before disconnecting the battery charger, otherwise an error message
will result (see page 55).

6.1 Connecting the power supply and battery charger

1
1) Slide the country-specific plug adapter onto the power supply until it locks into place (see fig. 1).
2) Connect the round, three-pin plug of the power supply to the receptacle on the battery charger (see fig. 2) so
that the plug locks into place.
INFORMATION: Ensure correct polarity (guide lug). Do not use force when connecting the cable plug
to the battery charger.
3) Plug the power supply unit into the outlet (see fig. 3).
→ The green LED on the back of the power supply lights up.
→ The LED ring (status indicator) on the rear of the charger lights up green to indicate the correct connection
to the power supply.
→ If the green LED on the power supply and the LED ring on the battery charger do not light up, there is an error
(see page 55).

6.2 Charging the prosthesis battery

INFORMATION
When the Protective Cover is installed, the battery charger cable has to point to the upper closure. A correct knee
joint charging process is only ensured with this alignment.