Preparing the product for use

6.2 AxonMaster

Various control programmes and switching methods as well as the option of making refined patient-specific set­
tings are available for optimum adaptation to the patient. The standard programmes can be selected using the
560X500=* AxonSoft adjustment software and include predefined control options and switching methods.
Two MyoBock electrodes, e.g., 13E200/202=*, or if necessary a pull switch, e.g., 9X18, can be connected to the
13E500 AxonMaster. For information please refer to the 647G590 instructions for use for the 13E500 AxonMaster
Input on AxonMaster
Input 1
Input 2
Input 3
► Remove the cover caps from the required inputs. The remaining inputs cannot be used at this time and there­
fore remain covered by the caps.

6.2.1 Controlling the Electronic Lock

The electronic lock can be controlled myoelectrically by the following means configured at the factory:
A long co-contraction on channels 1 and 2 of the AxonMaster releases or locks the lock.
A short co-contraction switches the gripping mode of the gripper.
A pulse on channel 3 of the AxonMaster releases or locks the lock.

6.2.2 Connecting the Electrodes

6.2.3 Installing the AxonMaster in the Upper Arm Socket

The AxonMaster can be secured with adhesive tape or a hook-and-loop fastening strap. The AxonMaster can be
covered with film to provide additional protection against moisture ingress.
Should the available space in the upper arm socket be too small an area for the AxonMaster (lamination dummy)
must be reserved during lamination of the inner socket.

6.2.4 Sealing the prosthesis

1) After laminating the lamination ring, remove the grease and any dirt.
2) Lubricate the groove, inner edge and thread of the lamination ring with 633F30 Special Grease.
10
Electrode
Closing the hand
Opening the hand
Pull switch (optional)
1) Remove the cover caps from the required inputs.
2) Seal the slots on the AxonMaster with silicone
grease.
3) Connect the electrode cables to the AxonMaster.
Observe the slot assignments:
Slot 1: close
Slot 2: open
Slot 3: pull switch (optional)