µBinary solvent manager
The μbinary solvent manager (μBSM) is a high-pressure pump that provides
steady (pulse-free) solvent flow at flow rates ranging from 0.0 to 6.0 μL/min
(under closed-loop control) or to 100.0 μL/min (under open-loop control) at
103,421 kPa (1034 bar, 15,000 psi).
The
See also:
the documentation CD, for more information about this module.
µSample manager - fixed loop
The μsample manager-fixed loop (μSM-FL) manages sample injection. It uses
a fixed loop-based mechanism to inject samples. After a puncture needle
pierces the covering of a well or vial, a sample needle extends through the
puncture needle and aspirates the sample. The aspirated sample is then
drawn toward the sample loop while weak-wash solvent is simultaneously
dispensed through the needle, behind the sample. The wash syringe then
pressurizes the needle from the tip, compressing the air gaps around the
sample and positioning the sample completely within the loop. When the
sample is in position, the injection valve is actuated to the inject position, and
high-pressure flow from the pump impels the sample onto the
chromatographic column.
The μSM-FL employs a thermal system that maintains a set temperature in
the sample compartment. When used in conjunction with the optional sample
organizer they create a combined, thermally controlled compartment.
The µSample Manager - Fixed Loop Overview and Maintenance
See also:
Guide on the documentation CD, for more information about this module.

Sample organizer

The optional sample organizer enables automatic processing and increased
throughput by storing multiple microtiter or vial plates and transferring them
to and from the μSM-FL.
The sample organizer's storage-shelf compartment can hold a selection of
ANSI plates. A heater/cooler thermally conditions the shelf compartment and
maintains the temperature at a specified set point.
The Sample Organizer Overview and Maintenance Guide, on the
See also:
documentation CD, for more information about this module.
μ
Binary Solvent Manager Overview and Maintenance Guide, on
March 10, 2014, 715003588 Rev. A
System components
19