Waters ACQUITY UPLC M-Class System Manual page 26

Hide thumbs Also See for ACQUITY UPLC M-Class:
Table of Contents

Advertisement

2 Optimizing Performance
Requirement:
Use high-quality solvents, buffers, and additives (LC/MS-grade,
prefiltered).
Use high-quality water (LC/MS-grade).
Follow these general recommendations when performing a microscale
analysis:
Always use solvent filters on tubing lines in solvent bottles.
Filter salt-containing buffers with a 0.2-μm filter membrane.
Keep concentrated stock solutions to use when you prepare working
solutions.
Make fresh buffer solutions daily.
Do not add fresh buffer to old buffer, which can promote microbial
growth.
Keep the seal-wash line primed.
Prime solvent lines during system startup.
Keep all solvent lines primed.
Do not block the degasser vent line. Trim the tubing, if necessary.
Eluent lines A1 and B1 do not pass through the degassers. When
Note:
working at flow rates less than 1μL/min, passing the eluent through the
degasser may cause depletion of any volatile mobile phase modifier. For
most applications, a Waters column packed with sub-2-micron particles
provides sufficient back pressure to retard outgassing. If outgassing
occurs. Waters recommends use of the optional solvent stabilization kit.
Do not submerge the waste or degasser vent lines in liquid.
See also:
Information for details on how to route waste and degasser tubing.
Flush buffers from the system, using water, immediately after you are
done with them.
Use 10 to 20% organic solvent in water as a storage solvent if you keep
the system idle for extended periods (longer than 24 hours).
If the system is idle for a day or two keep the system running rather
than shutting down.
26
March 10, 2014, 715003588 Rev. A
The
μ
Binary Solvent Manager Overview and Maintenance

Advertisement

Table of Contents
loading

Table of Contents