Isocratic Solvent Manager Splitter - Waters ACQUITY UPC2 System Manual

Hide thumbs Also See for ACQUITY UPC2:
Table of Contents

Advertisement

Figure 5–6:
1
2
50 inch x .010 inch ID
To the ccManager
µ
29.5 inch x 50
To the mass spectrometer
(Alternative: Remove this tubing to
connect a tool-free ECI (or APCI) probe
capillary from the mass spectrometer.)
Requirement:
The ABPR "health" check cannot be performed when a splitter is connected in the
solvent path. Temporarily isolate the flow cell's PDA-detector outlet from the inlet tee of the
ccManager.
5.1.3

Isocratic solvent manager splitter

You can adjust the amount of analytical flow diluted and diverted to the QDa detector. To do so, fit
the isocratic solvent manager (ISM) with a dual-detection or triple-detection flow-splitter kit.
See
also:
The ACQUITY QDa Detector Operator's Overview and Maintenance Guide for more
information on how to configure your QDa detector.
The ACQUITY Isocratic Solvent Manager Overview and Maintenance Guide for more
information on how to configure your ISM splitter.
MS splitter (SST)
m ID
November 16, 2016, 715004521 Rev. C
46 inch x .007 inch ID
From the CM-30S or PDA
4
3
67 inch x .004 inch ID
From the check valve outlet port
on the 515 makeup pump or, from
the ISM tee
Page 50