C Specifications; System Specifications; Quaternary Solvent Manager-R Specifications - Waters ACQUITY Arc System System Manual

Hide thumbs Also See for ACQUITY Arc System:
Table of Contents

Advertisement

C
Specifications
C.1 System specifications
The following table lists the system feature specifications.
Table C–1: System specifications
Feature
Dwell volume (total
system)
Operating flow rate range
Maximum operating
pressure
pH range
Unattended operation
Cycle time
C.2 Quaternary Solvent Manager-R specifications
Table C–2: QSM-R specifications
Attribute
Solvent capacity
Number of fluid
paths
Dwell volume
selection
Solvent conditioning
Specification
Path 1 = 1100 µL
Path 2 = 700 µL
0.001 to 5.000 mL/min, in 0.001 mL increments
9,500 psi
2 to 12
For leak sensors and safe leak handling: 96-hour
display of diagnostic data, via the console
<30 seconds, injection-to-injection
Specification
Total capacity of nine solvents with optional, integrated
solvent selection valve; blends as many as four solvents in
any combination (standard)
Three (Path 1, Path 2, and waste), with Arc Multi-flow path
technology (standard)
Automated with Arc Multi-flow path technology
Integrated vacuum degassing, four chambers