Waters ACQUITY UPLC M-Class System Manual page 85

Hide thumbs Also See for ACQUITY UPLC M-Class:
Table of Contents

Advertisement

µSM-FL performance specifications: (Continued)
Feature
Minimum sample required
Advanced sample manager
capabilities
Primary wetted materials
Sample organizer performance specifications:
Feature
Sample plate capacity
Maximum sample capacity
Sample compartment
temperature range
Temperature accuracy
Temperature stability
Features and performance specifications
Specification
3 μL residual, using Waters' total
recovery 2-mL vials (zero offset)
Load Ahead and Loop Offline mode,
valve cycling timed event, multi-load
with trapping independent of sample
loop volume
Fused silica, DLC, PEEK blend,
PEEK, Fluoropolymer, PPS,
fluoroelastomer, borosilicate glass,
PPS Blend, 316 stainless steel
Specification
Sample plate capacity is configured
based on the types and combinations
of plates being used:
• Maximum of 19 standard microtiter
plates, up to 15.5 mm high
• Maximum of 9 intermediate height
plates (or 2-mL vial holders), up to
40.0 mm high
• Maximum of 6 deep well plates (or
4-mL vial holders), up to 47.0 mm
high
Maximum of 7296 samples in nineteen
384-well plates
4.0 to 40.0 °C, settable to 0.1 °C
increments; maintains 19 °C below
ambient with a tolerance range
between -2 and +4 °C
±0.5 °C at sensor
±1.0 °C at sensor
March 10, 2014, 715003588 Rev. A
85