Column technology

Columns may be configured with the TVM using no valves for direct injection,
one or two valves for two-dimensional applications.
Columns
The TVM supports nanoscale (75-mm I.D.), capillary scale (300-mm I.D.) and
microscale (1.0-mm I.D.) separations. You can choose among these options
when selecting columns for the system:
Nanoscale analytical columns:
75 μm to 150 μm I.D.
50 mm to 250 mm length
Capillary-scale analytical columns:
300 μm I.D.
50 mm to 150 mm, length
Microscale analytical columns:
500 μm to 1 mm I.D.
50 mm to 150 mm length
Filter or guard column at 35 mm length
Trap columns:
180 μm to 500 μm I.D.
20 mm to 23 mm length
Ion exchange columns with SCX material
180 μm I.D.
240 mm length
The ACQUITY UPLC M-Class System documentation CD, for the
See also:
following documents:
The ACQUITY UPLC M-Class Trap Valve Manager Overview and
Maintenance Guide
The ACQUITY UPLC M-Class Columns Care and Use Manual
March 10, 2014, 715003588 Rev. A