Μsm-Fl Control Panel - Waters ACQUITY UPLC M-Class System Manual

Hide thumbs Also See for ACQUITY UPLC M-Class:
Table of Contents

Advertisement

3 Preparing the System
Solvent manager control panel: (Continued)
Control panel item
Stop flow
Warning indicator
You can access additional functions, shown in the following table, by
right-clicking anywhere in the solvent manager's control panel.
Additional functions in the solvent manager control panel:
Control panel function
Prime A/B solvents
Prime seal wash
Reset μBSM or ASM
Help
μSM-FL control panel
The μSM-FL control panel displays current sample-compartment
temperatures and set points. You can edit these values when the system is
idle by clicking the underlined value. You cannot edit threshold μSM-FL set
points while the system is running samples.
The cooling fans of the μSM-FL cease to circulate air whenever the
Tip:
sample compartment door is open. To ensure that the sample compartment
maintains a constant temperature, open its door only when
38
March 10, 2014, 715003588 Rev. A
Description
The Stop Flow control immediately stops all
flow from the solvent-manager.
Displays the warning indicator icon when the
liters pumped has been exceeded.
Description
Displays the Prime Solvents dialog box and
allows for manual changeover or refreshing
of solvent. Solvents are automatically
shunted to waste.
Starts priming the seal wash, which
lubricates the plungers, fills the tubing paths
with solvent, and flushes away solvent or
precipitated salts dragged past the plunger
seals from the high pressure side of the
piston chambers.
Resets the solvent manager following an
error condition.
Displays the instrument console's online
Help.
necessary.