Binary Solvent Manager - Waters ACQUITY UPLC I-Class IVD System Manual

Hide thumbs Also See for ACQUITY UPLC I-Class IVD:
Table of Contents

Advertisement

1 ACQUITY UPLC I-Class IVD System
ACQUITY UPLC I-Class Binary Solvent Manager (BSM)
ACQUITY UPLC I-Class Sample Manager - Flow-Through-Needle (SM -
FTN)
50 µL extension loop for SM-FTN
Column Heater with active pre-heating (CH-A) and tubing
Column Heater (CH-A) extension kit for mass spectrometer
Xevo TQ-S IVD mass spectrometer
ACQUITY UPLC I-Class System Startup kit, including solvent tray
module
Leak sensors
ACQUITY UPLC BEH C
ACQUITY UPLC I-Class IVD system documentation CD, part number
715003740IVD
Workstation and MassLynx, ACQUITY UPLC Console software, and
Connections INSIGHT software
FlexCart, optional

Binary solvent manager

The Waters ACQUITY UPLC I-Class IVD Binary Solvent Manager (BSM)
delivers solvent compositions for isocratic and binary gradient methods at
flow rates of between 0.01 and 2.0 mL per minute. Its features include in-line
filters upstream of a primary check valve, the Waters
2
Valve (i
Valve), automated priming functions, and daily system-setup
routines.
The design of the binary solvent manager is optimized for sub-2-µm particle
liquid chromatography. The upper pressure limit is set at 124,106 kPa
(1241 bar, 18,000 psi) per pump. A maximum flow rate of 1 mL/min is
permitted at system backpressures to 124,106 kPa (1241 bar, 18,000 psi),
which falls within the range of optimal linear velocity for sub-2-µm particle
columns of ID 1.0 mm to 3.0 mm.
The ACQUITY UPLC I-Class IVD Binary Solvent Manager
See also:
Overview and Maintenance Guide (part number 715003739IVD) for more
details.
18
July 7, 2014, 715003736IVD Rev. C
column, 1.7 µm, 2.1 x 50 mm, IVD
18
®
Intelligent Intake

Advertisement

Table of Contents
loading