B Specifications
µSM-FL performance specifications:
Feature
Injection volume range
Injection linearity
Injection modes
Sample manager precision
Number of sample plates
Maximum sample capacity
Sample compartment
temperature range
Temperature accuracy
Temperature stability
Injection needle wash
84
March 10, 2014, 715003588 Rev. A
Specification
0.1 μL to 100.0 μL in 0.1-μL
increments; 5-μL loop standard with 1,
2, 5, 10, 20, 50 and 100- μL optional
loops; unlimited injection volume
during trapping
>0.999 (default needle) from 20% to
80% volume for the 5-μL loop
Full loop mode for optimal
quantitation and dispersion; Partial
loop mode for fastest cycle time and
sample volume flexibility
<1% area RSD 0.2 to 1.9-μL injection;
<0.5% area RSD 2.0 to
10.0-μL-injection
Any two of the following:
• 96 and 384-well microtiter plates
• 48 position 2.00-mL vial plates
• 48 position 0.65-mL
micro-centrifuge tube plates
• 24 position 1.50-mL centrifuge tube
plates
768 in two 384-well plates, or 96 in
2-mL vial holders, plus 4 additional
positions for dilution functions or
standards
4.0 to 40.0 °C, settable in 0.1 °C
increments; maintains 19.0 °C below
ambient with a tolerance range
between -2.0 and +4.0 °C.
±0.5 °C at sensor
±1.0 °C at sensor
Integrated, active, programmable,
dual wash