6. Settings

6.1.System settings
A SYSTEM setting window appears when you select [Setup] - [System] from the file
menu. A system setting file can be changed on the SYSTEM setting window. All spectral
measured value data items are deleted when you open the SYSTEM setting window.
Therefore, open the SYSTEM setting window after saving the required spectral measured
value data. Enter a password before opening the SYSTEM setting window. Enter the
initial value "uspm" as the password in this case.
Menu
TARGET
COM CTLR
STORE SIZE
MIN WAVE
PV8959-F4E005
Select the equipment to be connected. For a near-infrared
micro-spectrophotometer (USPM-RU-W), set this to
USPM-RU-W.
Select the connection ID number of an internal arithmetic
unit. Usually, set this to [0001]. This item does not require
change.
Sets the total number of spectral measured value data items
saved in storage; a maximum of 20 data items can be
secured in storage. Usually, set this to 20. The setting range
is an integer value of 1 to 20.
Sets the lower-limit value of the wavelength to be displayed
and analyzed; data files may not be readable when you
change this value. The setting value varies depending on the
equipment to be connected. For a near-infrared
23
Description