Olympus USPM-RU-W Series Operation Manual page 77

Optical measuring instrument
Hide thumbs Also See for USPM-RU-W Series:
Table of Contents

Advertisement

MAX WAVE
LUMINANCE BIT
PIXEL NUMBER
LUMINANCE LVL
FOCUS XY RANGE
CT XY RANGE
CALIBRATION
ND
BACKGROUND 2
PV8959-F4E005
micro-spectrophotometer (USPM-RU-W), set this to 380.
Sets the upper-limit value of the wavelength to be displayed
and analyzed; data files may not be readable when you
change this value. The setting value varies depending on the
equipment to be connected. For a near-infrared
micro-spectrophotometer (USPM-RU-W), set this to 1050.
Sets the number of data bits for the spectral light intensity;
usually, set this to 12. This item does not require change.
The maximum value of spectral light intensity changes when
this item is changed.
Sets the number of pixels in a photosensor; the setting value
varies depending on the equipment to be connected. For a
near-infrared micro-spectrophotometer (USPM-RU-W), set
this to 1024.
Sets the lower-limit value for the spectral light intensity; enter
a ratio for the maximum spectral light intensity value as a
percentage. For less than the lower-limit value set here,
[UNDER] changes to red in the status display of the
operation panel. The sampling time adjusted by [Adjust
sampling] varies depending on this item. Usually, set this to
80. The setting range is an integer value of 0 to 100.
Sets the gravity tolerance of a FOCUS window; the gravity
position display color of the FOCUS window turns red when
this item lapses out of the set range. Usually, set this to 1.
The setting range is an integer value of 0 to 100.
Sets the gravity tolerance of a CT window; the gravity
position display color of the CT window turns red when this
item lapses out of the set range. Usually, set this to 10. The
setting range is an integer value of 0 to 100.
Sets the movement amount in the direction of the
wavelength; enter negative numbers when moving this in the
long wavelength direction. Enter positive numbers when
moving this in the short wavelength direction. Usually, set
this to 0. Set this also according to the wavelength calibration
procedure.
Sets a double background function; a BACKGROUND2
button is displayed on the operation panel of the main
window when this item is set to "Enable." This enables
background measurement. Usually, set this to "Disable."
24