7.3.[LIVE] window
FOCUS and CT images are displayed on the [LIVE] window. They can be switched by
selecting [FOCUS] or [CT] from [Image] from the file menu or double-clicking on the screen.
The current image can be saved when you click the right button of the mouse to select [SAVE].
JPEG or BMP can be selected as the saving format. The current image can be copied to the
clipboard when you select [COPY].
Menu
Display image
name
Center-of-gravity
coordinates
Exposure time
setting
In-focus state
Luminance graph
PV8959-F4E005
In-focus state
Cente-of-gravity coordinates
Indicates the displayed image; [FOCUS] is
displayed when the current display is a FOCUS
camera image. [CT] is displayed when it is a CT
camera image.
Displays the center-of-gravity coordinate position;
this item is displayed in red when the
center-of-gravity coordinate position is out of the
range set using a system setting file.
Sets the exposure time; the current exposure time
is displayed in the numeric value on the left.
Automatic adjustment is properly performed when
you click the AUTO check box. "Wait...." is
displayed while automatic adjustment is in
progress.
Displays the in-focus state; the position in which a
value is lowest near the in-focus state is an
in-focus position.
Displays the luminance state
Display image name
Exposure time setting
Description