Olympus USPM-RU-W Series Operation Manual page 25

Optical measuring instrument
Hide thumbs Also See for USPM-RU-W Series:
Table of Contents

Advertisement

(8)
Pull out the gas spring from the hole.
(9)
Insert a new gas spring into the hole.
(10) Push the replacement jig into the gas spring access hole to compress the gas spring.
(11) Insert and fix hinge pin 1 into the upper edge of the new gas spring with a slotted
screwdriver.
(12) Remove the replacement jig of the gas spring.
(13) Fix the covers with the screws.
(14) Turn the load adjuster counterclockwise to loosen it.
(1), (2)
Load adjuster
(5), (6), (7)
PV8344F4E013
Photometric head
Four screws on the
back
Upper cover
Jig for replacing
the gas spring
Gas spring
access hole
Hinge pin 1
(9) - (14) is the reverse procedure of (1) - (8).
23
(3), (4)
(8)
Two screws on the
top side of the rack
Jig for replacing
the gas spring
Gas spring