Olympus USPM-RU-W Series Operation Manual page 47

Optical measuring instrument
Hide thumbs Also See for USPM-RU-W Series:
Table of Contents

Advertisement

Failure in the
measurement of
wavelength direction
Display of wavelength
in LUMINANCE
window is incorrect
Display of wavelength
in GRAPH window is
incorrect
Computer's
misbehaving
operation.
Failure in
measurement data.
PV8344F4E013
Check the correctness of reference setting.
(11)
Check whether sampling interval is appropriate.
(12)
Check for injection of ambient light from fluorescent lamps or
(13)
the sun.
Refer to 5.3 Wavelength offset to conduct a offset.
(1)
Turn off the controller supply unit and reboot the computer.
(1)
Turn off the controller supply unit and reboot the computer.
(1)
Plug off and reinsert USB cable, then reboot the software.
(1)
Turn off the Control Power Suppy Box, then reboot the
(2)
computer.
45

Advertisement

Table of Contents
loading

This manual is also suitable for:

Uspm-w-b