The Information window opens that displays the sampling interval. After that, Click
OK to end the operation.Then, the shutter operates again.
4.9. Obtaining the background value
(1)
Click BACKGROUND1 on the operation panel.
A clicky sound is generated, and then a warning alarm sounds as the process
finishes.
As the data is saved, an asterisk (*) appears at the left of BACKGROUND1 in the
operation panel.
4.10. Obtaining the reference value
(1)
Click REFERENCE on the operation panel.
The shutter sounds, the reference value is captured. After that, the shutter operates
again, and a warning alarm sounds as the process finishes.
As the data is saved, an asterisk (*) appears at the left of REFERENCE in the
operation panel.
4.11. Spectroscopy
(1)
Click MEASURE on the operation panel.
The shutter sounds, indicating the start of spectroscopy. When alarm sounds and
spectroscopy is ended, a shutter sounds again.
The result is displayed in the sub-windows. Up to 20 results can be stored.
If you measure film thickness, refer to 4.12 Analysis of film thickness. If you
measure color differences, refer to 4.13 Calculation of color difference.
4.12. Analysis of film thickness
(1)
Open the LAYER settings window.
Choose [Setup]-[Layer] on the File menu. The LAYER window opens.
File menu