Olympus USPM-RU-W Series Operation Manual page 38

Optical measuring instrument
Hide thumbs Also See for USPM-RU-W Series:
Table of Contents

Advertisement

Item
AUTO
UNIT
USE REFLECTANCE
OF SUBSTANCE
RANGE
METHOD
LAYER
(3)
Click the THICKNESS button to conduct an analysis of film thickness.
4.13. Calculation of color difference
Calculation of color difference is extracted from the values of spectroscopy.
For detail, refer to 5.3.1 Calculation of color difference in the USPM-SA Software
Operation Manual.
(1)
Open the COLOR DIFFERENCE window.
Choose [Analyze]-[Color Difference] on the File menu. The COLOR DIFFERENCE
window opens.
File menu
(2)
Use the < > buttons to select a DATA NO for the calculation of color difference in
each pane of DATA1 and DATA2. Color difference is automatically calculated as the
numbers are defined.
PV8344F4E013
If enabled, analysis of film thickness is automatically done after the easurement.
Unit of the film thickness displayed in the analysis.
Any spectroscopy data file (*.csv/*.dat) is available as reflectance of substance. The first
data in the data file will be used as the reflectance of substance. Details on reflectance of
substance will be provided later.
Wavelength range to be analyzed in the analysis of film thickness. High accuracy can be
achieved by restricting the range to the wavelength range of high reliance of spectral
analysis.
The method of analysis of film thickness. [PV] denotes peak-valley method, [FFT]
frequency analysis method, and [FIT] curve fit method. For detail on each method, refer to
6.3 Layer setting in the UPSM-SA Software Operation Manual.
Design parameter of the layer to be analyzed. There is no need to set up if METHOD is
[PV] and UNIT is [OPTICAL]. Otherwise, it must be properly configured. For details on the
setting, refer to 6.3.1 Layer setting in the UPSM-SA Software Operation Manual.
36