Checklist For Inspection Of The Lighting Device (Lamp House) - Olympus USPM-RU-W Series Operation Manual

Optical measuring instrument
Hide thumbs Also See for USPM-RU-W Series:
Table of Contents

Advertisement

Checklist for inspection of the lighting device (lamp house)

Read the manuals carefully before conducting an inspection.
Inspections on a regular basis (at least once every half year and when replacing a lamp)
is recommended in order to ensure safe use.
Inspect the items in the list, and put a checkmark next to each passed item and a circle
next to each item having some problem.
If there is a circle next to any item examined in the latest inspection, stop using the unit
and ask your supplier to inspect the unit or to replace it with a new lighting device.
If you find any failure in any other items than those on the list or some kind of failure in
other components of our products aside from the lighting device, stop using the product
and ask your supplier to inspect the unit.
Repairs, replacement and inspections out of warranty are available for a fee.
Contact your supplier with any questions you may have.
1.
Eight years have elapsed since purchase, or the total time of lamp use has
exceeded 20,000 hours.
2.
Sometimes lamp does not illuminate even when you turn on the main switch
3.
The light blinks if you move the lamp cable or the body itself.
4.
The lamp cable or another part is abnormally hot.
5.
There is an odor like something is burning when in use.
6.
Light flickers even after the lamp is replaced.
7.
A deformation, unsteadiness or loose part is discovered when installing the
lighting device (e.g., a lamp house is too tight to insert).
8.
The color of any terminal or the lamp socket is drastically changed. Or, the color is
different between the left one and the right one.
9.
There is change of color, deformation or cracks in the lamp house.
10. There is melting, cracks, deformation or solidification in the lamp cable or wiring.
11. Frequency of repair is higher than before for the same type of other units brought
into service around the same time.
*
Make photocopies of the list if needed.
PV8344F4E013
Inspection checklist
Terminal (levers)
48
Result
(date of inspection)
/
/
/
/
Lamp socket

Advertisement