B Specifications; Features And Performance Specifications - Waters ACQUITY UPLC M-Class System Manual

Hide thumbs Also See for ACQUITY UPLC M-Class:
Table of Contents

Advertisement

B
Specifications
The specifications presented in this document depend on the conditions that exist in individual
laboratories. Refer to the site preparation guide or contact Waters Technical Service for additional
information about specifications.
B.1 Features and performance specifications
The following tables list the performance specifications for the ACQUITY UPLC M-Class system
and system components.
Table B–1: System specifications
Feature
Flow control
Integrated leak management
Operating flow rate range
System synchronization
Maximum operating pressure
pH range
Total system delay volume
Unattended operation
Solvent storage
Specification
Direct, non-split, and automatic solvent flow
control algorithms provide pulse-free flow
Safe leak handling; system diagnostic
capability to identify and pinpoint system leaks
at nL/min scale
0.0 nL/min to 100.0 µL/min without flow
splitting
Injection synchronization between pumps and
the sample manager enhances retention time
reproducibility
103,421 kPa (1034 bar, 15,000 psi)
2 to 10
<1 µL
Full 96-hour diagnostic data display through
console software
Solvent tray accommodates eluents for up to
three wash solvents and one seal wash solvent
November 26, 2019, 715003588 Revision C