Table B–5: Sample organizer performance specifications
Feature
Sample plate capacity
Maximum sample capacity
Sample compartment temperature range
Temperature accuracy
Temperature stability
Table B–6: TVM performance specifications
Feature
Column capacity
Switching valves
Column compartment temperature range
Column compartment temperature accuracy
Column compartment temperature stability
Primary wetted materials
Specification
Sample plate capacity is configured based on
the types and combinations of plates being
used:
Maximum of 7296 samples in nineteen 384-
well plates
4.0 to 40.0 °C, settable to 0.1 °C increments;
maintains 19 °C below ambient with a
tolerance range between -2 and +4 °C
±0.5 °C at sensor
±1.0 °C at sensor
Specification
Single column, 75 µm to 4.6 mm I.D. and up to
250 mm in length
Two six-port, two-position valves with
independent programmable control
Ambient +9.0 to 90.0 °C, settable to 0.1 °C
increments
(Column compartment temperature lower
range is equal to ambient +5° C)
±0.5 °C at sensor
±0.3 °C at sensor
DLC, PEEK blend, and 316 stainless steel
November 26, 2019, 715003588 Revision C
Page 90
• Maximum of 19 standard microtiter plates,
up to 15.5 mm high
• Maximum of nine intermediate height plates
(or 2-mL vial holders), up to 40.0 mm high
• Maximum of six deep well plates (or 4-mL
vial holders), up to 47.0 mm high