Tighten the fitting finger-tight plus as much as an additional 1/6-turn using a 1/4-inch open-end
wrench.
Reinstalled tightening
2.6.1.13 V-detail fitting (for use with small, fused silica capillary tubing)
First use
Short flats
Compression screw
Two-piece stainless steel ferrule
Tighten the fitting finger-tight plus an additional 3/4-turn using a 1/4-inch open-end wrench.
Tip:
To prevent band spreading, ensure that the tubing is fully bottomed in the fitting before
tightening the compression screw.