Note:
If ZenFit capillary tubing is used in the system, then you must install the M to ZenFit
adapter fitting in each ASM and BSM outlet.
Tighten the fitting finger-tight plus an additional 1/2-turn using a 1/4-inch open-end wrench.
First-use tightening
Reinstalled
Flats
Compression screw
One-piece stainless steel ferrule
Tighten the fitting finger-tight plus as much as an additional 1/6-turn using a 1/4-inch open-end
wrench.
Reinstalled tightening