Table B–4: µSM-FL performance specifications
Feature
Injection volume range
Injection linearity
Injection modes
Sample manager precision
Number of sample plates
Maximum sample capacity
Sample compartment temperature range
Temperature accuracy
Temperature stability
Injection needle wash
Minimum sample required
Advanced sample manager capabilities
Primary wetted materials
Specification
0.1 µL to 100.0 µL in 0.1-µL increments; 5-µL
loop standard with 1, 2, 5, 10, 20, 50, and 100-
µL optional loops; unlimited injection volume
during trapping
>0.999 (default needle) from 20% to 80%
volume for the 5-µL loop
Full loop mode for optimal quantitation and
dispersion; partial loop mode for fastest cycle
time and sample volume flexibility
<1% area RSD 0.2 to 1.9-µL injection; <0.5%
area RSD 2.0 to 10.0-µL-injection
Any two of the following:
768 in two 384-well plates, or 96 in 2-mL vial
holders, plus four additional positions for
dilution functions or standards
4.0 to 40.0 °C, settable in 0.1 °C increments;
maintains 19.0 °C below ambient with a
tolerance range between -2.0 and +4.0 °C
±0.5 °C at sensor
±1.0 °C at sensor
Integrated, active, programmable, dual wash
3 µL residual, using Waters total recovery vials
(zero offset)
Load Ahead and Loop Offline mode, valve
cycling timed event, multi-load with trapping
independent of sample loop volume
Fused silica, DLC, PEEK blend, PEEK,
fluoropolymer, PPS, fluoroelastomer,
borosilicate glass, PPS Blend, 316 stainless
steel
November 26, 2019, 715003588 Revision C
Page 89
• 96- and 384-well microtiter plates
• 48-position 2.00-mL vial plates
• 48-position 0.65-mL micro-centrifuge tube
plates
• 24-position 1.50-mL centrifuge tube plates