Automatic Exposure Control
Nominal shortest irradiation time
The shortest exposure duration that may occur in an AEC sequence is:
Nominal Shortest Duration Time:
Allowable AEC deviation
Deviation Range of Current Time Product:
CAUTION
XGB6-12-1014, Rev 1.04
For detailed information of the loading factors of each automatic
modes, please read respective sections describing these modes.
Chapter 7 - Fluoro Imaging Modes
1.2 sec
≤±(10% + 0.2 mAs)