Half Dose Fluoroscopy

Activating the mode: DR Fluoroscopy
To switch to this mode:
STEP
Ranges of loading factors
The ranges of the loading factor values are listed below:
DR Fluoroscopy
kV
mA
7.7
Half Dose Fluoroscopy
In Half Dose Fluoroscopy mode, the G-Arm system sets the loading factors
such as kV and mA values automatically by simply reducing the mA values
of the Automatic Continuous Fluoroscopy or Automatic Pulse Fluoroscopy
by half.
During fluoroscopic imaging, the X-ray is continuously/intermittently gener-
ated, depending on the fluoroscopy mode that has been previously activat-
ed.
Activating the mode: Half Dose Fluoroscopy
To switch to this mode:
STEP
Ranges of loading factors
The ranges of the loading factor values are listed below:
½ Dose Auto Cont. Fluoroscopy
kV
mA
½ Dose Automatic Pulse Fluoroscopy (Normal)
kV
XGB6-12-1014, Rev 1.04
In any mode, press the
The LCD screens on the control panel display the current mode -
"PULSE | DR FLUO".
40~120 kV
40.0 mA
In any mode, press the
The LCD screens on the control panel display the current mode -
"CONTINUOUS | HALF AUTO FLUO" or "PULSE | HALF AUTO FLUO".
40~120 kV
0.2~2.0 mA
40~120 kV
Chapter 7 - Fluoro Imaging Modes
key.
key.
75