Chapter 7 - Fluoro Imaging Modes
Activating the Tissue or Bone feature
To activate this feature:
STEP
Tissue/Bone Parameters
The parameters used for the Tissue and Bone features are listed below:
Tissue
Bone
7.10

Automatic Exposure Control

In automatic modes, if you release the footswitch or the
trol panel immediately after a press, the G-Arm system will initiate the AEC
(Automatic Exposure Control) sequence, in which the loading factors (kV,
mA, and exposure duration) are automatically controlled according to the
physical characteristics of the patients.
Automatic modes, where the AEC sequence may take over the control, in-
cludes the following:
Automatic Continuous Fluoroscopy mode
Automatic Pulse Fluoroscopy mode
Half Dose Automatic Continuous Fluoroscopy mode
Half Dose Automatic Pulse Fluoroscopy mode
DR Fluoroscopy mode
Activate the AEC sequence
Take the following procedure to invoke the AEC sequence:
STEP
78
In any mode, press the Tissue
The LCD screens on the control panel display the currently selected
feature - capitalized characters "T" or "B" - in the lower right corners.
Note: You can neither activate nor deactivate both of these two fea-
tures at the same time. The Bone feature is automatically activated
upon system start.
Default Brightness
Default Contrast
1
0
Press and immediately release either of the following:
any of the footswitch pedals
either of the
keys on the control panel
Automatic Exposure Control
or Bone
key.
2
1
key on the con-
B6 User Manual