System; Output Electric Power; X-Ray Tube Assembly - Whale MultiScan G-Arm System XGB6-12-1026 User Manual

Table of Contents

Advertisement

System

10.1
System
Resolution:
Low contrast resolution:
Brightness ladder level:
Total equivalent filtration:
permanent filtration
additional filtration (irremovable) 2.0 mmAl (at 80 kV)
Distortion rate:
10.2

Output Electric Power

When performing real-time manual pulse fluoroscopy from both AP and LT
planes, the G-Arm system achieves the highest power, as indicated below:
Nominal electric power:
Highest electric power:
10.3

X-ray Tube Assembly

Major parameters of the X-ray tube assembly configuration are listed as be-
low.
Tube Assembly Model:
Tube Model:
Maximum Voltage to X-ray Tube:
Focal spots:
Target angle:
Anode:
Anode material:
Minimum inherent filtration:
Thermal capacity:
Weight:
Leakage Radiation:
Additional Filtration:
For detailed specifications on X-ray tube and tube assembly, consult the fol-
lowing accompanying documents:
B6
Specifications of X-ray Tube (Doc#: XGB6-12-00573 Ve r: 1.0)
XGB6-12-1014, Rev 1.04
Chapter 10 - Technical Specifications
≥1.6 Lp/mm
≤ 2%
≥ 10 degrees
4.0 mmAl (at 80 kV)
2.0 mmAl (at 80 kV)
≤ 5%
2.3 kW (100 kV, 15 mA, 1 s)
1.8 kW (120 kV, 15 mA)(bi-plane imaging,
peak), 1.2 kW (120 kV, 15 mA) (bi-plane imag-
ing, maximum average)
E-40R HF IMD/XR05
IMD/XR05
120 kV
0.3/0.6 mm
10°
Rotating Anode
Tungsten
2.0 mmAl
600 kJ
19.6 kg
< 1 mGy/h at 120 kV and 4 mA
2.0 mmAl