Chapter 13 - Electromagnetic Compatibility Statement
# Emission Test
2
Harmonic emissions IEC 61000-3-2
3
Voltage fluctuations/flicker emis-
sions IEC 61000-3-3
Electromagnetic Environment – Guidance (#1 in the above table)
The G-Arm System uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
13.3

Electromagnetic Immunity

# IMMUNITY Test
1
Electrostatic discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
2
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
3
Surge
IEC 61000-4-5
±1 kV Line to line
±2 kV Line to earth
4
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
5
Radiated RF
IEC 61000-4-3
Electromagnetic Environment - Guidance
Electrostatic discharge (ESD) (#1 in the above table)
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst (#2 in the above table)
Mains power quality should be that of a typical commercial or hospital envi-
ronment.
Surge (#3 in the above table)
Mains power quality should be that of a typical commercial or hospital envi-
ronment.
154
Electromagnetic Immunity
Compliance
N/A
N/A
Compliance Level
±2 kV, ±4 kV, ±6 kV contact
±2 kV, ±4 kV, ±8 kV air discharge
±2 kV for power supply lines
±1 kV Line to line
±2 kV Line to earth
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
B6 User Manual