Chapter 7 - Fluoro Imaging Modes
Body part
Limbs
Cervical spine
Thoracic spine/lumbar/sacral
Hip
7.2

Automatic Continuous Fluoroscopy

In Automatic Continuous Fluoroscopy mode, the G-Arm system sets the
loading factors such as kV and mA values automatically. During fluoroscopic
imaging, the X-ray is continuously generated.
In this mode, the mA and the kV values alter to achieve optimal image qual-
ity based on the particular conditions of the patient.
Activating the mode: Automatic Continuous Fluoroscopy
To switch to this mode:
STEP
Ranges of loading factors
The ranges of the loading factor values are listed below:
Automatic Continuous Fluoroscopy
kV
mA
7.3

Manual Continuous Fluoroscopy

In Manual Continuous Fluoroscopy mode, you manually set the loading fac-
tors such as kV and mA values according to the particular conditions of dif-
ferent patients. During fluoroscopic imaging, the X-ray is continuously
generated.
Activating the mode: Manual Continuous Fluoroscopy
To switch to this mode:
STEP
72
Recommended mode
Continuous Fluoroscopy
Continuous Fluoroscopy
Pulse Fluoroscopy
Pulse Fluoroscopy
In any mode, press the
The LCD screens on the control panel display the current mode -
"CONTINUOUS | AUTO FLUO".
40~120 kV
0.2~ 4.0 mA
In the Automatic Continuous Fluoroscopy mode, press either of the
kV or mA adjustment keys (
Continuous Fluoroscopy mode.
Automatic Continuous Fluoroscopy
key.
or
) to switch to the Manual
B6 User Manual