Chapter 9 - Maintenance
STEP 3
STEP 4
For the test to pass, the measured kV values shall be within the range of kV
values (±10%), as indicated in the following table:
Table 9-1: Range of Measured kV Values (Manual Continuous Fluoroscopy)
kV
40
72
108
120
Table 9-2: Range of Measured kV Values (Manual Pulse Fluoroscopy)
kV
40
72
108
120
40
72
108
120
9.10
Cleaning and Disinfection
WARNING
Clean the equipment by using cotton cloth and do not use abrasive de-
tergent, organic solvent or any kind of detergent containing any kind
of organic solvent, or aggressive detergent containing alcohol.
Do not use spray or spray disinfectant as it can penetrate the surface of
the equipment and damage the internal electric elements.
Use sterile drapes to cover the G-Arm and footswitch to protect the
equipment from blood spills and to create a sterilized environment for
surgery.
132
Initiate fluoroscopic imaging according to Table 9-1 (p138) and re-
cord the measured kV values displayed on dose meter.
Switch to Manual Pulse Fluoroscopy mode, repeat step 3.
mA
kV Range
2
36 ~ 44
2
64.8 ~ 79.2
2
97.2 ~ 118.8
2
108 ~ 132
mA
kV Range
15
36 ~ 44
15
64.8 ~ 79.2
15
97.2 ~ 118.8
15
108 ~ 132
30
36 ~ 44
30
64.8 ~ 79.2
30
97.2 ~ 118.8
30
108 ~ 132
Before cleaning the G-Arm system, always disconnect the AC power
cable from the AC power receptacle. Do NOT allow any liquids to drip
or splash on the equipment, especially the electric components, to
avoid hazards of short circuit, electric shock and fire.
Cleaning and Disinfection
B6 User Manual