Touch-screen Controls

5.3
Touch-screen Controls
The control monitor provides a touch-screen for interacting with the Work-
station software. Using the touch-screen (and the keyboard for text-entry)
you can:
Enter and view patient information
View and process images/videos
Set imaging parameters
Annotate, compare, zoom, or rotate images
Print images
Generate dose summary reports
For detailed information, please refer to the chapter "Workstation Software"
(p87).
5.4

Collimator

The collimator is a fully automatic, microprocessor controlled device de-
signed to provide accurate X-ray collimation. The collimator automatically
adjusts itself upon SID (source to image distance) change. The collimator can
be manually adjusted through corresponding collimator keys on the control
panel.
The collimator will automatically adjust its aperture to control the X-ray field
according to the current SID value.
XGB6-12-1014, Rev 1.04
Figure 5-7: Control Unit Monitors
1
Viewing monitors
2
Control monitor with touch-screen
3
Control panel with two LCD screens
4
Keyboard drawer