Turning The System On - Whale MultiScan G-Arm System XGB6-12-1026 User Manual

Table of Contents

Advertisement

Chapter 2 - Start-up
2.2

Turning the System On

WARNING
CAUTION
To turn the system on:
STEP 1
STEP 2
STEP 3
26
Figure 2-5: Turning the system on
1
Emergency stop
2
Circuit Breaker
At initial setup or when removing from storage, stabilize the equip-
ment at ambient temperature and humidity for 4 hours before apply-
ing power. Failure to observe this warning might damage the
equipment.
If the equipment has not been used for an extended amount of time
(over 30 days), turn it on and leave it idle without taking any exposures
for 2~3 hours in order to maintain the required vacuum in the image
intensifier.
When the G-Arm system is in the Off state, the G-Arm will not respond
to any control commands even with the circuit breaker being switched
on.
Turn the emergency stop (1) clockwise. The button will be released
if it has been previously activated.
Flip up the circuit breaker (2).
Press the Power On button
begins the power up sequence.
Turning the System On
above the control panel. The system
B6 User Manual