Chapter 8 - Workstation Software
AP (Single):
LT (Single):
Bi-plane:
Total:
PCT (%):
C/Counter:
C/Time (s):
P/Counter:
P/Time (s):
C/DAP (uGy*m^2):
C/Dose (mGy):
P/DAP (uGy*m^2):
P/Dose (mGy):
To generate a dose summary report:
STEP 1
STEP 2
STEP 3
106
Exposure(s) conducted on the AP plane alone.
Exposure(s) conducted on the LT plane alone.
Exposure(s) conducted simultaneously on both AP and LT
planes.
The sum of respective values for all exposures taken on AP,
LT and Bi-plane.
Percentage that the continuous/pulse mode accounts in
total exposure count, total exposure duration, total DAP, or
total dosage.
Count of the exposures taken in the automatic continuous
fluoroscopy mode.
Duration of the exposures taken in the automatic continu-
ous fluoroscopy mode.
Count of the exposures taken in the automatic pulse fluo-
roscopy mode.
Duration of the exposures taken in the automatic pulse flu-
oroscopy mode.
Total DAP of the exposures taken in the automatic continu-
ous fluoroscopy mode.
Total dosage of the exposures taken in the automatic con-
tinuous fluoroscopy mode.
Total DAP of the exposures taken in the automatic pulse
fluoroscopy mode.
Total dosage of the exposures taken in the automatic pulse
fluoroscopy mode.
Create or open a study, make sure it contains fluoroscopic images.
Press
to show more buttons, and then press Dose Sum-
mary Report
.
A dose summary report about this study is generated.
On the Dose Summary Report screen, you can continue to perform
the following operations:
Saving the report as a picture: Press Generate Image to save
the report as a bitmap (*.bmp) file.
Saving the report as a PDF file: Press PDF to print the report in
the PDF format.
Printing the report: Press Print to print the report on A4-size
paper.
Generating Dose Reports
B6 User Manual