Automatic Pulse Fluoroscopy

Ranges of loading factors
The ranges of the loading factor values and the step values it will take when
you manually adjust the loading factors are listed below:
Manual Continuous Fluoroscopy
kV
mA
7.4
Automatic Pulse Fluoroscopy
In Automatic Pulse Fluoroscopy mode, the G-Arm system sets the loading
factors such as kV and mA values automatically. During fluoroscopic imag-
ing, the X-ray is intermittently generated.
In this mode, the mA and the kV values alter to achieve optimal image qual-
ity based on the particular conditions of the patient.
Note: Simultaneous bi-plane (AP & LT) fluoroscopy imaging, namely, taking ex-
posures from both AP and LT planes, can only be initiated from the Auto/Manual
Pulse Fluoroscopy mode. While in other continuous fluoroscopy modes, if the
AP/LT foot pedal (blue & yellow colored) is pressed, the system automatically
switches to the pulse modes.
Activating the mode: Automatic Pulse Fluoroscopy
To switch to this mode:
STEP
Ranges of loading factors
The ranges of the loading factor values are listed below:
Automatic Pulse Fluoroscopy
kV
mA
XGB6-12-1014, Rev 1.04
The LCD screens on the control panel display the current mode -
"CONTINUOUS | MANUAL FLUO".
40~120 kV
Step 1 kV
0.2~4.0 mA
Step 0.1 mA
In any mode, press the
The LCD screens on the control panel display the current mode -
"PULSE | AUTO FLUO".
40~120 kV
0.2~30 mA
(1-plane imaging)
0.2~30 mA
(2-plane imaging)
Chapter 7 - Fluoro Imaging Modes
key.
73