Appendix IV - QC Test Log Form
A(IV)-3 Accuracy Test of kV Values
Table IV-1: kV Accuracy Test in Manual Cont. Fluo. Mode
kV
40
60
80
100
110
120
Table IV-2: kV Accuracy Test in Manual Pulse Fluo. Mode
kV
40
60
80
100
110
120
178
Criteria
Value
36~44
54~66
72~88
90~110
99~121
108~132
Criteria
Value
36~44
54~66
72~88
90~110
99~121
108~132
Accuracy Test of kV Values
Pass?
Time
Pass?
Time
Signature
Signature
B6 User Manual